DS360 Ultra Low Distortion Function Generator Revision 2.4 (11/2013)
Operating Manual and
Programming Reference
Model DS360
Ultra Low Distortion
Function Generator
1290-D Reamwood Avenue
Sunnyvale, CA 94089 U.S.A.
Phone: (408) 744-9040, Fax: (408) 744-9049
email: info@thinkSRS.com • www.thinkSRS.com
Copyright © 1999, 2007, 2008, 2011, 2012, 2013 by SRS
All Rights Reserved
Przeglądanie stron 0
1 2 3 4 5 6 ... 139 140

Podsumowanie treści

Strona 1 - Function Generator

DS360 Ultra Low Distortion Function Generator Revision 2.4 (11/2013) Operating Manual and Programming Reference Model DS360 Ultra Low Distortion

Strona 2

viii Specifications DS360 Ultra Low Distortion Function Generator Two-Tone Type Sine-Sine, Sine-Square Sine Frequency 0.01 Hz to 200.000 kHz S

Strona 3 - WARNING!

5-18 Performance Tests DS360 Ultra Low Distortion Function Generator 8. Bursts These tests measures the accuracy of the burst levels. Setup Connect th

Strona 4

Performance Tests 5-19 DS360 Ultra Low Distortion Function Generator 9. DC Offset These tests measure the offset accuracy of the function of the

Strona 5 - Table of Contents

5-20 Performance Tests DS360 Ultra Low Distortion Function Generator d) Measure the output voltage on the DVM. Record the results. 4) Measure the

Strona 6 - Chapter 6 Circuitry

Performance Tests 5-21 DS360 Ultra Low Distortion Function Generator 10. Output Impedance These tests confirms the output source resistors of th

Strona 7 - Table of Figures

5-22 Performance Tests DS360 Ultra Low Distortion Function Generator 7) Disconnect the + BNC output from the DVM. Connect the - BNC output to the DVM.

Strona 8

Performance Tests 5-23 DS360 Ultra Low Distortion Function Generator DS360 Performance Test Record Page 1 of 4 Serial Number ___________

Strona 9 - Specifications

5-24 Performance Tests DS360 Ultra Low Distortion Function Generator DS360 Performance Test Record Page 2 of 4 4. Amplitude Accuracy (Con

Strona 10 - Amplitude

Performance Tests 5-25 DS360 Ultra Low Distortion Function Generator DS360 Performance Test Record Page 3 of 4 5. Harmonic Distorti

Strona 11

5-26 Performance Tests DS360 Ultra Low Distortion Function Generator DS360 Performance Test Record Page 4 of 4 9. DC Offset Offset Linea

Strona 12 - General

Circuit Description 6-1 DS360 Ultra Low Distortion Function Generator Chapter 6 Circuit Description This chapter provides descriptions of the

Strona 13 - Abridged Command List

Specifications ix DS360 Ultra Low Distortion Function Generator Noise Broadband Noise (for a 1 kHz sine wave into a high impedance). 40 Vpp

Strona 14 - Status Reporting Commands

Circuit Description DS360 Ultra Low Distortion Function Generator 6-2

Strona 15 - Getting Started

Circuit Description DS360 Ultra Low Distortion Function Generator 6-3 Overview The DS360’s circuitry is divided into two major sections: the dig

Strona 16 - 1-2 Getting Started

Circuit Description DS360 Ultra Low Distortion Function Generator 6-4 Microprocessor and System Clocks U401 is a Z80 general purpose microprocessor.

Strona 17 - Front Panel Operation

Circuit Description DS360 Ultra Low Distortion Function Generator 6-5 Front Panel Control The front panel display board consists of eight 7 segm

Strona 18 - Continuous Waveforms

Circuit Description DS360 Ultra Low Distortion Function Generator 6-6 Analog Board This section covers schematics ANA-1 through ANA-7 at the end of t

Strona 19 - Frequency Sweeps

Circuit Description DS360 Ultra Low Distortion Function Generator 6-7 signal. Each filter is a second order, state variable, low pass filter. J3

Strona 20 - Tone Bursts

Circuit Description DS360 Ultra Low Distortion Function Generator 6-8 U503). Q503, Q504 and Q506 are configured as a cascode, to maintain a constant

Strona 21 - Chapter 2

Circuit Description 6-9 DS360 Ultra Low Distortion Function Generator Front Panel This section covers schematic FP-1 and FP-2 at the end of this

Strona 22 - 2-2 Basics

Circuit Description DS360 Ultra Low Distortion Function Generator 6-10

Strona 23 - Traditional Generators

Parts List DS360 Ultra Low Distortion Function Generator 6-11Digital Board Parts List This section covers schematics DIG-1 through DIG-7. Digital B

Strona 24 - 2-4 Basics

x Specifications DS360 Ultra Low Distortion Function Generator General Computer Interface GPIB and RS-232 standard. All instrument functions can be

Strona 25 - Front Panel Features

Parts List DS360 Ultra Low Distortion Function Generator 6-12C 722 5-00225-548 .1U AXIAL Capacitor, Ceramic, 50V,+80/-20% Z5U AX C 723 5

Strona 26 - Keypad

Parts List DS360 Ultra Low Distortion Function Generator 6-13N 104 4-00334-425 10KX5 Resistor Network SIP 1/4W 2% (Common) N 501

Strona 27

Parts List DS360 Ultra Low Distortion Function Generator 6-14U 118 3-00049-340 74HC74 Integrated Circuit (Thru-hole Pkg) U 119 3

Strona 28 - Outputs

Parts List DS360 Ultra Low Distortion Function Generator 6-15)Analog Board Parts List This section covers schematics ANA-1 through ANA-7. Analog B

Strona 29 - Indicators

Parts List DS360 Ultra Low Distortion Function Generator 6-16C 237 5-00225-548 .1U AXIAL Capacitor, Ceramic, 50V,+80/-20% Z5U AX C 238 5

Strona 30 - Parameter LEDs

Parts List DS360 Ultra Low Distortion Function Generator 6-17C 403 5-00012-501 330P Capacitor, Ceramic Disc, 50V, 10%, SL C 404

Strona 31 - Display

Parts List DS360 Ultra Low Distortion Function Generator 6-18C 703 5-00321-509 2200U Capacitor, Electrolytic, 50V, 20%, Rad C 704 5

Strona 32 - Rear Panel Features

Parts List DS360 Ultra Low Distortion Function Generator 6-19C 760 5-00100-517 2.2U Capacitor, Tantalum, 35V, 20%, Rad C 761

Strona 33

Parts List DS360 Ultra Low Distortion Function Generator 6-20K 404 3-00523-335 G6AK-234P-ST-UC Relay K 405 3

Strona 34 - 2-14 Basics

Parts List DS360 Ultra Low Distortion Function Generator 6-21Q 507 3-00022-325 2N3906 Transistor, TO-92 Package Q 508

Strona 35 - Operation

Abridged Command List xi DS360 Ultra Low Distortion Function Generator Abridged Command List Syntax Commands which have a question mark in pa

Strona 36 - 3-2 Operation

Parts List DS360 Ultra Low Distortion Function Generator 6-22R 225 4-00161-407 2.49K Resistor, Metal Film, 1/8W, 1%, 50PPM R 226 4

Strona 37 - Power On

Parts List DS360 Ultra Low Distortion Function Generator 6-23R 418 4-00135-407 1.50K Resistor, Metal Film, 1/8W, 1%, 50PPM R 419

Strona 38 - Setting Functions

Parts List DS360 Ultra Low Distortion Function Generator 6-24R 541 4-00347-407 7.50K Resistor, Metal Film, 1/8W, 1%, 50PPM R 542 4

Strona 39 - Operation 3-5

Parts List DS360 Ultra Low Distortion Function Generator 6-25R 621 4-00522-407 243 Resistor, Metal Film, 1/8W, 1%, 50PPM R 622 4-00490-407

Strona 40 - Relative Amplitude

Parts List DS360 Ultra Low Distortion Function Generator 6-26U 210 3-00094-340 LM311 Integrated Circuit (Thru-hole Pkg) U 301 3

Strona 41 - Operation 3-7

Parts List DS360 Ultra Low Distortion Function Generator 6-27Front Panel Parts List This section covers schematics FP-1 through FP-2. Front Panel

Strona 42

Parts List DS360 Ultra Low Distortion Function Generator 6-28D 138 3-00012-306 GREEN LED, Rectangular D 139 3

Strona 43 - Output Configuration

Parts List DS360 Ultra Low Distortion Function Generator 6-29Z 0 0-00048-011 6-32 KEP Nut, Kep Z 0 0-0

Strona 44 - 3-10 Operation

Parts List DS360 Ultra Low Distortion Function Generator 6-30Z 0 6-00119-614 FT82-77 Iron Powder Core Z 0 6-001

Strona 45 - Modify Function

Parts List DS360 Ultra Low Distortion Function Generator 6-31Distortion Filter Parts List This section covers schematics PROGR-1. Distortion Filte

Strona 46 - Sweep Rate

xii Abridged Command List DS360 Ultra Low Distortion Function Generator Setup Control Commands *IDN? 4-17 Returns the DS360 device identification s

Strona 47 - Operation 3-13

Parts List DS360 Ultra Low Distortion Function Generator 6-32Miscellaneous and Chassis Parts List Miscellaneous and Chassis Assembly Ref No. Ref No

Strona 48 - Sine and Square Wave Bursts

Getting Started 1-1 DS360 Ultra Low Distortion Function Generator Chapter 1 Getting Started These examples are designed to acquaint the first ti

Strona 49 - BurstCount

1-2 Getting Started DS360 Ultra Low Distortion Function Generator

Strona 50 - Bandwidth

Getting Started 1-3 DS360 Ultra Low Distortion Function Generator Front Panel Operation Parameters are set in the DS360 using the front panel

Strona 51 - Operation 3-17

Getting Started 1-4 DS360 Ultra Low Distortion Function Generator Continuous Waveforms This section discribes how to set up different continuous

Strona 52 - Instrument Setup

Getting Started 1-5 DS360 Ultra Low Distortion Function Generator Frequency Sweeps This section discribes how to set up a linear or logarithmi

Strona 53 - Operation 3-19

Certification Stanford Research Systems certifies that this product met its published specifications at the time of shipment. Stanford Research System

Strona 54 - Self-Tests

1-6 Getting Started DS360 Ultra Low Distortion Function Generator Tone Bursts This section discribes how to set up tone bursts. The DS360 can produce

Strona 55 - Troubleshooting

Basics 2-1 DS360 Ultra Low Distortion Function Generator Chapter 2 Basics In this Chapter Introduction 2-3 Traditional Function Generators

Strona 56 - Operational Errors

2-2 Basics DS360 Ultra Low Distortion Function Generator

Strona 57 - Operation 3-23

Basics 2-3 DS360 Ultra Low Distortion Function Generator Introduction to Precision Waveform Synthesis The DS360 uses Direct Digital Synthesis an

Strona 58 - 3-24 Operation

2-4 Basics DS360 Ultra Low Distortion Function Generator DDS with Advanced Signal Processing A block diagram for the DS360 is shown in Figure 1. The D

Strona 59 - Remote Programming

Basics 2-5 DS360 Ultra Low Distortion Function Generator Front Panel Features –+WHT NOISEDIGITALBALUNBALLIN SWPSHIFTON/STBY40V max.DEPTHTRIG SR

Strona 60 - 4-2 Remote Programming

2-6 Basics DS360 Ultra Low Distortion Function Generator Keypad Figure 2-3 Keypad Function Keys These keys control the main function output. The [∧]

Strona 61 - Index of Commands

Basics 2-7 DS360 Ultra Low Distortion Function Generator change causes the currently displayed modify function to become invalid, the display w

Strona 62 - 4-4 Remote Programming

2-8 Basics DS360 Ultra Low Distortion Function Generator Outputs –+40V max.+–COMMON40 VDC MAX.GROUNDTTLTTLTRIGGER/GATE INSYNC OUTXLR OUTPUT BNC OUTPUT

Strona 63 - Introduction

Basics 2-9 DS360 Ultra Low Distortion Function Generator Indicators REM SRQACT ERRSTATUSFREQAMPLSTP/BWSTR/CTRIGRCOUNTOFFSETOUTPUTONOFFTRIG&apos

Strona 64 - Interface Ready and Status

i DS360 Ultra Low Distortion Function Generator Safety and Preparation for Use WARNING! Dangerous voltages, capable of causing injury or death,

Strona 65 - Remote Programming 4-7

2-10 Basics DS360 Ultra Low Distortion Function Generator Parameter LEDs These LEDs indicate which parameter is currently displayed in the parameter d

Strona 66 - Command Syntax

Basics 2-11 DS360 Ultra Low Distortion Function Generator Display Hz dBkHz dBV% dBmVp-pVrmsVDC Figure 2-6 Display Parameter Display This 8 dig

Strona 67 - Remote Programming 4-9

2-12 Basics DS360 Ultra Low Distortion Function Generator Rear Panel Features FUSEPULLRS232 (DCE)SWEEP OUT BURST TRIG OUTIEEE-488 STD PORTCHASSISG

Strona 68 - 4-10 Remote Programming

Basics 2-13 DS360 Ultra Low Distortion Function Generator Outputs and Computer Interface Figure 2-8 Outputs & Computer Interfaces IEEE-

Strona 69 - Remote Programming 4-11

2-14 Basics DS360 Ultra Low Distortion Function Generator

Strona 70 - Modify Function Commands

Operation 3-1 DS360 Ultra Low Distortion Function Generator Chapter 3 Operation The following sections describe the operation of the DS360. Th

Strona 71 - Remote Programming 4-13

3-2 Operation DS360 Ultra Low Distortion Function Generator Troubleshooting 3-21 GPIB Problems 3-21 RS232 Problems 3-21 Error Messages

Strona 72 - 4-14 Remote Programming

Operation 3-3 DS360 Ultra Low Distortion Function Generator Power On Power On When power is first applied to the DS360 the unit will display its

Strona 73 - Setup Control commands

3-4 Operation DS360 Ultra Low Distortion Function Generator Setting Functions The following section describes how to set the parameters for the basic

Strona 74

Operation 3-5 DS360 Ultra Low Distortion Function Generator If the function is set to White Noise or Pink Noise the character of the noise is f

Strona 75 - Remote Programming 4-17

ii DS360 Ultra Low Distortion Function Generator

Strona 76 - 4-18 Remote Programming

3-6 Operation DS360 Ultra Low Distortion Function Generator 0001121000. ≤≤AmpToneAmpTone If the amplitude of either Tone1 or Tone 2 is set outside o

Strona 77 - Status Byte Definitions

Operation 3-7 DS360 Ultra Low Distortion Function Generator Unbalanced 600 ΩΩΩΩ Output Amplitude Ranges Function Vpp Vrms dBm dBV Sine 5 µV -

Strona 78 - 4-20 Remote Programming

3-8 Operation DS360 Ultra Low Distortion Function Generator * The rms, dBm and dBV values for White & Pink Noise are based on the total power i

Strona 79 - Example Programs

Operation 3-9 DS360 Ultra Low Distortion Function Generator Output Configuration This section describes how to configure the output type and sou

Strona 80 - 4-22 Remote Programming

3-10 Operation DS360 Ultra Low Distortion Function Generator The output impedance changes for unbalanced and balanced. Each output (positive and negat

Strona 81 - Remote Programming 4-23

Operation 3-11 DS360 Ultra Low Distortion Function Generator Modify Function This section describes how to set the modify functions on the DS360

Strona 82 - 4-24 Remote Programming

3-12 Operation DS360 Ultra Low Distortion Function Generator minimize this, keep stop frequencies below 130 kHz when sweeping at maximum amplitude. Sw

Strona 83 - Performance Tests

Operation 3-13 DS360 Ultra Low Distortion Function Generator (from the TRIGGER IN, computer interface or front panel) and the actual beginning o

Strona 84 - 5-2 Performance Tests

3-14 Operation DS360 Ultra Low Distortion Function Generator computer interfaces. In addition, the output can be gated on and off from the TRIGGER/GAT

Strona 85 - Getting Ready

Operation 3-15 DS360 Ultra Low Distortion Function Generator If a value that is greater than or equal to the current burst rate is entered the u

Strona 86 - If a Test Fails

Table of Contents iii DS360 Ultra Low Distortion Function Generator Table of Contents Safety and Preparation For Use i Table of Contents i

Strona 87 - Filter Component Values

3-16 Operation DS360 Ultra Low Distortion Function Generator the current value. If an out of range value is entered, the unit will beep, display “Dept

Strona 88 - 5-6 Performance Tests

Operation 3-17 DS360 Ultra Low Distortion Function Generator Center Frequency The center frequency can vary from 0 Hz to 200.000 kHz in 200 Hz i

Strona 89 - 1. Front Panel Test

3-18 Operation DS360 Ultra Low Distortion Function Generator Instrument Setup This section describes the DS360’s default settings, storing and recalli

Strona 90 - 2. Self Tests

Operation 3-19 DS360 Ultra Low Distortion Function Generator message “Store Done” will be displayed to indicate that the settings have been sto

Strona 91 - 3. Frequency

3-20 Operation DS360 Ultra Low Distortion Function Generator DS360’s input data queue. The data is displayed in ASCII hex format, where each character

Strona 92 - 4. Amplitude

Operation 3-21 DS360 Ultra Low Distortion Function Generator Troubleshooting Nothing Happens on Power On Make sure that the power entry module o

Strona 93 - Performance Tests 5-11

3-22 Operation DS360 Ultra Low Distortion Function Generator value.The DS360 is configured to send or receive: 2 stop bits, 8 data bits and no parity

Strona 94 - 5-12 Performance Tests

Operation 3-23 DS360 Ultra Low Distortion Function Generator Self-Test Errors Message Meaning CPU ERR The DS360 detected a problem with the CP

Strona 95 - Procedure

3-24 Operation DS360 Ultra Low Distortion Function Generator LD DONE This message occurs after succesfully loading the DSP memory. RECALL Recall m

Strona 96 - 5-14 Performance Tests

Remote Programming 4-1 DS360 Ultra Low Distortion Function Generator Chapter 4 Remote Programming In this Chapter Index of Commands 4-2 Introd

Strona 97 - 6. Waveforms

iv Table of Contents DS360 Ultra Low Distortion Function Generator Burst Test 5-18 DC Offset Test 5-19 Output Impedance Test 5-21 DS360 Performa

Strona 98 - 5-16 Performance Tests

4-2 Remote Programming DS360 Ultra Low Distortion Function Generator Status Reporting Commands 4-16 Hardware Test & Calibration Commands 4-17

Strona 99 - 7. Sweeps

Remote Programming 4-3 DS360 Ultra Low Distortion Function Generator Index of Commands Variables i, j, k, n integers x real numbers Function O

Strona 100 - 8. Bursts

4-4 Remote Programming DS360 Ultra Low Distortion Function Generator Setup Control Commands *IDN? 4-17 Returns the DS360 device identification strin

Strona 101 - 9. DC Offset

Remote Programming 4-5 DS360 Ultra Low Distortion Function Generator Introduction The DS360 Ultra Low Distortion Function Generator may be remo

Strona 102 - 5-20 Performance Tests

4-6 Remote Programming DS360 Ultra Low Distortion Function Generator Command Format Communication with the DS360 uses ASCII characters. Commands may b

Strona 103 - 10. Output Impedance

Remote Programming 4-7 DS360 Ultra Low Distortion Function Generator When using the GPIB interface, serial polling may be used to check the No

Strona 104 - 5-22 Performance Tests

4-8 Remote Programming DS360 Ultra Low Distortion Function Generator Command Syntax The four letter mnemonic in each command sequence specif

Strona 105

Remote Programming 4-9 DS360 Ultra Low Distortion Function Generator Function Output Control Commands Note: The analog output control commands

Strona 106 - Page 2 of 4

4-10 Remote Programming DS360 Ultra Low Distortion Function Generator OUTE (?) i The OUTE command disables the output for i=0 and enables the output f

Strona 107 - Page 3 of 4

Remote Programming 4-11 DS360 Ultra Low Distortion Function Generator The TTAA? query will return the amplitude in the currently displayed unit

Strona 108 - Page 4 of 4

Table of Contents v DS360 Ultra Low Distortion Function Generator Table of Figures Figure 2-1 Block Diagram 2-4 Figure 2-2 Front Panel

Strona 109 - Circuit Description

4-12 Remote Programming DS360 Ultra Low Distortion Function Generator Modify Function Commands Note: Most of the modify function parameters may be sel

Strona 110 - Circuit Description

Remote Programming 4-13 DS360 Ultra Low Distortion Function Generator i Source 0 INTERNAL 1 EXTERNAL 2 SINGLE 3 GATE (burst only) If the trigg

Strona 111 - Digital Board

4-14 Remote Programming DS360 Ultra Low Distortion Function Generator NRCT (?) x The NRCT command sets the Noise Rate Count to x. The NRCT? query retu

Strona 112

Remote Programming 4-15 DS360 Ultra Low Distortion Function Generator Setup Control commands *IDN? The *IDN? common query returns the DS360 dev

Strona 113 - Circuit Description

4-16 Remote Programming DS360 Ultra Low Distortion Function Generator Status Reporting Commands Note: See tables at the end of the programming section

Strona 114 - Analog Board

Remote Programming 4-17 DS360 Ultra Low Distortion Function Generator Hardware Test and Calibration Commands *TST? The *TST? common query runs

Strona 115

4-18 Remote Programming DS360 Ultra Low Distortion Function Generator $PST (?) i The $PST command sets the DS360 post-amplifier attenuators to range i

Strona 116 - Power Supply

Remote Programming 4-19 DS360 Ultra Low Distortion Function Generator Status Byte Definitions The DS360 reports on its status by means of three

Strona 117 - Programmable Resistor Board

4-20 Remote Programming DS360 Ultra Low Distortion Function Generator Standard Event Status Byte Bit Name Set When 0 Unused 1 Unused 2 Query Err

Strona 118

Remote Programming 4-21 DS360 Ultra Low Distortion Function Generator Example Programs Using Microsoft C with the National Instruments GPIB car

Strona 119 - Digital Board Parts List

vi Contents DS360 Ultra Low Distortion Function Generator

Strona 120 - Parts List

4-22 Remote Programming DS360 Ultra Low Distortion Function Generator Example1: GPIB Communication in C language /* C program to demonstrate commun

Strona 121 - Parts List

Remote Programming 4-23 DS360 Ultra Low Distortion Function Generator Example 2: RS232 communication in BASIC language BASIC program to demo

Strona 122

4-24 Remote Programming DS360 Ultra Low Distortion Function Generator

Strona 123 - Analog Board Parts List

5-1 DS360 Ultra Low Distortion Function Generator Chapter 5 Performance Tests The performance tests described in this section are designed t

Strona 124

5-2 Performance Tests DS360 Ultra Low Distortion Function Generator

Strona 125

Performance Tests 5-3 DS360 Ultra Low Distortion Function Generator Getting Ready Parameters are set in the DS360 using the front panel keypad o

Strona 126

5-4 Performance Tests DS360 Ultra Low Distortion Function Generator Instrument Critical Specification Recommended Model Analog Oscilloscope 100 MH

Strona 127

Performance Tests 5-5 DS360 Ultra Low Distortion Function Generator Distortion Measurements The notch filters described here are used with a spe

Strona 128

5-6 Performance Tests DS360 Ultra Low Distortion Function Generator Total harmonic distortion can be calculated from the following formula:

Strona 129

Performance Tests 5-7 DS360 Ultra Low Distortion Function Generator 1. Front Panel Test This test verifies the functionality of the front panel

Strona 130

Specifications vii DS360 Ultra Low Distortion Function Generator Specifications Waveforms Sine Frequency 0.01 Hz to 200.000 kHz THD 1 VRMS U

Strona 131

5-8 Performance Tests DS360 Ultra Low Distortion Function Generator 2. Self Tests This test verifies that the DS360’s memory and processors are functi

Strona 132

Performance Tests 5-9 DS360 Ultra Low Distortion Function Generator 3. Frequency This test measures the frequency accuracy of the DS360. Setup C

Strona 133

5-10 Performance Tests DS360 Ultra Low Distortion Function Generator 4. Amplitude This test measures the amplitude accuracy of the DS360. For the sin

Strona 134

Performance Tests 5-11 DS360 Ultra Low Distortion Function Generator b) Press [∨] (output down) to select balanced output. c) Press

Strona 135 - Front Panel Parts List

5-12 Performance Tests DS360 Ultra Low Distortion Function Generator c) Record the VRMS value from the DVM and the VPEAK-PEAK value from the scope.

Strona 136

Performance Tests 5-13 DS360 Ultra Low Distortion Function Generator 5. Harmonic Distortion and Noise These tests check the harmonic distortion

Strona 137

5-14 Performance Tests DS360 Ultra Low Distortion Function Generator a) Connect the + output of the DS360 directly to the spectrum analyzers input.

Strona 138

Performance Tests 5-15 DS360 Ultra Low Distortion Function Generator 6. Waveforms These procedures check the various output waveform characteris

Strona 139 - Distortion Filter Parts List

5-16 Performance Tests DS360 Ultra Low Distortion Function Generator b) Press [∨] (function down) to select 2-Tone. [FREQ] [1] [kHz] [AMPL]

Strona 140

Performance Tests 5-17 DS360 Ultra Low Distortion Function Generator 7. Sweeps These procedures test the flatness of frequency sweeps. Setup Con